Throughput | - Up to 133 EMIT® tests per hour with two reagents
- Up to 65 EMIT® tests per hour with three reagents
|
Reagent system | - EMIT reagent rotor — with 13 positions for 14 mL bottles and 13 positions for 28 mL bottles
- All positions in both rotors can be assigned as R1, R2, and R3
- The reagent rotor compartment is cooled to 8–12°C (absolute up to 25°C ambient temperature)
- Reagent volumes:
– Reagent 1 volume 110–399 μL – Reagent 2 volume 10–289 μL – Reagent 3 volume 10–289 μL - Aspirated reagents are heated, with level detector and integrated stirrer
|
Sample system – providing semi-continuous loading | - Sample rotor contains:
– 50 positions for barcoded tubes – 12 positions for tubes without barcodes – 1 blank and 1 washing position
- Primary tubes (13 or 16 mm OD)
- Positions can contain 5 mL or 10 mL primary tubes or sample cups
- Sample volume 1 – 30 µL per test, programmable in steps of 0.1 µL
- Sample probe with level detection and integrated stirrer
|
Pipetting system | - Flex Fluidics syringes
- Reagent syringe 1000 μL
- Sample syringe 100 μL
|
Cuvette rotor | - Semi-disposable rotor with 48 cuvettes, path length 6.8 mm
- Minimum measuring volume 200 μL
- Measuring temperature 37°C controlled by Peltier elements
|
Washing units | - Cuvette rinsing with 4 x 500 μL of treated water:
– The unit is equipped with liquid sensors – Waste is separated into diluted and concentrated waste
– Cuvettes are dried before use |
Light source | Quartz-iodine lamp 12 V – 20 W |
Wavelength range | - Automatic wavelength selection by an 8-position filterwheel (340, 415, 505, 546, 570, 600, 660, 700 nm)
- Half bandwidth 8 to 12 nm
|
Photometric range | Absorbance -0.1 to 3.0 |
Ambient temperature | - 15–32°
- Relative humidity 15–85% (non condensing)
|
Calculation modes | - Linear regression
- Modified cubic spline
- Syva® 4 parameter logit log
|
Quality control | - Up to 6 controls per parameter (up to 12 controls programmable per rotor configuration)
- Westgard rules
- Levey-Jennings plots
|
Computer | - CPU: Intel Celeron M 575 2 GHz
- Touchscreen monitor: 15.6 inch, 1366 x 768 pixels
- Ports: 1 serial, 4 USB, and 1 ethernet port
|
Printer | - The application supports one printer
- Most printers supported by WINDOWS can be connected
|
Barcode reader (option) | The Viva-ProE™ is equipped with an internal barcode reader to read barcoded sample tubes. An optional hand-held barcode reader is available. |
Languages | English, French, German, Italian, Spanish; others on request |
Dimensions | Width: 125 cm/49.2 in Depth: 62 cm/24.4 in Height, excl monitor: 75 cm/29.5 in Weight: approx. 93 kg/205 lbs (excl. monitory arm and panel PC) |