Eclipse RD Cyclotron

Eclipse RD Cyclotron
 
Contact Us

General Specifications

  • 40 µA per target
  • Typical performance 2,000 mCi 18F in 120 min per target
  • Typical performance 4,000 mCi duals
  • 13N ammonia 100 mCi/10 min
  • 15O gas 2,000 mCi/10 min
  • 11C CO2 1,500 mCi/50 min
  • 18F F2 400 mCi/60 min
  • 11C specific activity is >25 Ci/µA
  • 2 mR/hour radiation field at room boundary (22ft. x 23ft.) dual 40s
  • Single-vacuum window
  • Helium and water cooling required
  • Does not support solid targets


Areas of Research:

  • Small-volume targets with high production yields, up to 8 targets; 40 µA
  • Medical diagnostics (pre-clinical animal testing)
  • Research and Clinical Systems (drug development)
  • First-time cyclotron or replacement cyclotron implementation


Typical System Performance:

  • Unique 4- and 8-position carousel capable of running 40 µA
  • Energy: 11 MeV
  • Magnet: 4-sector azimuthally varying field, single-coil strip

     ° Power: 3kW
     ° Mean Field: 1.2 T
     ° Valley to Hill Gap Ratio: 27:1
     ° Magnetic Field @Room Boundary: <1 Gauss

  • RF System

     ° Fundamental Mode: 72MHz
     ° Max. Amplifier Power: 10kW

  • Ion Source: Penning Ion Gauge

     ° Max. H-current: 2mA

  • Beam Extraction: Carbon Foils

     ° # foils/carousel: 3
     ° Foil Lifetime: 10,000 µA-hr
     ° Beam Size: 8mm (RD)

  • Vacuum (in Torr)

     ° Base pressure (mm Hg) 5x10-7
     ° Operating pressure (mm Hg) <1x10-5

  • Power Consumption

     ° Standby <7kW
     ° Operating 35 kW

  • Heat Rejection

     ° Air 52,000 BTU/hr
     ° Water 90,000 BTU/hr

  • Shielding

     ° RD 2.0 mR/hr (20µSv/hr) total n+g @ 2x40 µA at room boundary



Operating Characteristics:

  • Operating temperature: 70° F (21° C)
  • Operating humidity: 40-55% non-condensing R.H.
  • Control Band: ± 1° C (3° F)


Electrical / Power Requirements:

  • USA / Asia: 208 (± 5%) VAC; 150A, 3-phase 60Hz standard (50Hz optional); 380 (± 5%) VAC; 150A, 3phase 60Hz (optional)
  • Europe: 208 (± 5%) VAC; 150A, 3-phase 50Hz standard Agency approvals:
  • ISO 9001 and ISO 13485 by BSI (British Standards Institute)


Dimensions:

Cyclotron

  • Width: 59 in. (150 cm)
  • Depth: 59 in. (150 cm)
  • Height: 68 in. (173 cm)
  • Weight: 22,000 lbs (10000 kg)


Movable Shields

  • Width: 95 in. (240 cm)
  • Depth: 73 in. (184 cm)
  • Height: 90 in. (230 cm)
  • Weight: 14,450 lbs (31800 kg)


Stationary Shields

  • Width: 26 in. (66 cm)
  • Depth: 73 in. (184 cm)
  • Height: 90 in. (230 cm)
  • Weight: 7,400 lbs (3,400 kg)


Control Cabinets (1 unit)

  • Width: 32 in. (81 cm)
  • Depth: 56 in. (142 cm)
  • Height: 83 in. (210 cm)
  • Weight: 2,200lbs (1,000 kg)


Heat Exchange Cabinet

  • Width: 32 in. (81 cm)
  • Depth: 32 in. (81 cm)
  • Height: 83 in. (210 cm)
  • Weight: 1,000 lbs (450 kg)